Microprocessor Flame Photometer
Product Details :
Application | Industrial Use |
Brand | MANTI |
Automation Grade | Automatic |
Product Type | Biochemical Analysis |
We have carved out a reputed position in manufacturing, trading, exporting and supplying high quality Microprocessor Flame Photometer. Our offered Photometer is extensively used for the testing of sodium, potassium, calcium and lithium. Under the strict supervision of our skilled supervision, this photometer is manufactured using latest micro controller technology and advanced engineering techniques so as to give enhanced result and reproducibility. Our provided photometer has soft touch membrane key for easy operation.
Key Features Are :
- Microprocessor Based with Printer Interface
- Auto Gas Cutoff
- Auto Flame Failure Detection
- Four Measurements in Single Aspiration
- Sample Data Storage Facility
- Curve Calibration Programming Using Maximum of 5 Standards
- Multiple Calibration Curve Saving Facility
- Data Processing Through Curve Fitting
- Techniques, Least Square, Quadratic
- Direct Results in ppm and meq
- Useful for Medical Soil Testing
Technical Specifications :
Mode | MT-124 | Serum | Urine |
Range | Na: 0.1-100 ppmUpto 250 meq/l, 1:100 dilK: 0.1-100 ppmUpto 250 meq/l, 1:100 dilCa: 15-100 ppmUpto 250 meq/l, 1:100 dilLi: 0.5-100 ppmUpto 250 meq/l, 1:100 dil | Na: 0-2001:100 dilK: 0-101:100 dilCa: - Li: 0-21:5 dil | Na: 0-2501:100 dilK: 0-1001:100 dilCa: - Li: 0-101:2 dil |
Sensitivity | General Mode: Na:0.5 ppm, K:0.5 ppm, Li:0.5 ppm, Ca:15 ppm | ||
Filters | Na K Supplied (Ca Li are optional at extra cost) | ||
Resolution | 0.1 ppm/meq | ||
Reproducibility | 2% FS, Â 2 Digits | ||
Curve Fit Accuracy | 2% FS | ||
Display | 20 x 4 Alphanumeric backlit LCD display | ||
Average Time | In-built in Software | ||
Flame System | LPG dry oil free air | ||
Detector | Silicon Photodiode | ||
Calibration | Upto 5-Point Calibration with curve fitting software | ||
Gas Control | Adjustable with knobs | ||
Ignition System | Auto Ignition | ||
Automiser | Axial flow type | ||
Printer | Provision for attachment of Dot-Matrix Printer of HP Laserjet/Inkjet Printer (non-win printers) with centronics parallel interface | ||
RS-232C Interface | Available (PC Software at extra cost) | ||
Power | 230 V 10% AC, 50 Hz | ||
Dimensions | 365 x 245 x 220 mm (L x B x H) (Approx.) | ||
Weight | Main Unit : 7.5 Kg. (Approx.) | ||
Compressor Unit | |||
Air Supply | Oil free mini compressor unit with pressure regulator | ||
Combustion Gas LPG | Controlled by precision regulator | ||
Power | 230 V 10% AC, 50 Hz | ||
Dimensions | 290 x 255 x 210 mm (L x B x H) (Approx.) | ||
Weight | 7.5 Kg (Approx.) |