Click Image for Gallery
Tag Flash Point Tester:
TAG-4
- Tag closed-cup flash point tester (ASTM D56 and more)
- Flash point applications of jet fuels, solvents, and more
- Two cooling systems, liquid or air, with Peltier elements
- Two ignition systems, electric and gas, included with each unit
The automatic Tag closed-cup tester measures the flash point, the lowest temperature at which the vapors of a sample ignite upon the application of an ignition source. The tester can also be used to detect the ability of a sample to yield flammable vapors when maintained at an equilibrium temperature by the flash/no flash method.
TAG 4 is suitable for flash point applications of jet fuels, solvents, chemicals, etc.
Key features
Benefits at a Glance
- Two cooling systems with Peltier elements are available: an economical air-cooled model and a low-temperature version with external water cooling
- The two ignition systems in each unit give you the flexibility to work with an electric or gas igniter: if the gas ignition is used, the electric igniter will automatically relight the gas flame during the test, if necessary
- The One-Twist swivel-around multi-function head guarantees the safe and easy one-hand connection of all sensors and actuators such as the shutter release, igniter and detector contacts
- An extended measuring range up to 110 °C enables you to go beyond the conventional scope of the method
Convenient Operation
- Test initiation by pre-installed standard test methods and ability to customize methods by user-defined programs
- Outside flame sensors increase the system’s safety, especially for work with samples that have no conventional flash point but burn outside the test cup (e.g. due to flame-inhibiting components)
- Splashproof touch-key panel with large LCD
- A multi-detector combines the flash point detector and the temperature probe in a solid housing with a perfectly adjusted immersion depth
Customized User Flexibility
- PC-software FPPNet (optional) for convenient remote handling and test data storage
- Insulation hood (optional) to protect the cover and shutter assembly from icing during low-temperature flash point tests
- Draft deflector (optional) recommended for gas ignition tests
- Calibration set (optional)
- Small-volume sample cup (optional)
Technical specifications :
Application range (°C/°F selectable) | 10 °C to 110 °C (air-cooled) -30 °C to 110 °C (liquid-cooled) |
---|---|
Ignition | Gas and electric (interchangeable) |
Heating | According to method or user-defined |
Cooling of Peltier elements | Air-cooled (by built-in fan) Liquid-cooled (with tap water or circulation cooler) |
Flash detection | Thermocouple |
Barometric pressure correction | Automatic correction |
Safety | Overheat protection, automatic shut-off |
Interfaces | 2x RS232 |
Gas supply | 50 mbar of propane or butane |
Power supply | 115 V/230 V, 50 Hz/60 Hz, 180 W (air-cooled) 230 V, 50 Hz/60 Hz or 115 V, 60 Hz, 150 W (liquid-cooled) |
Dimensions | 230 mm x 470 mm x 470 mm (W x D x H) |
Weight | 8 kg |
Resolution | 0.1 °C |