Click Image for Gallery
Dissolution Testers
Product Details :
- Enables filtration, pre-heating, deaeration, and precise dosing of media in the vessel within the dissolution bath
- 21 ltr media tank capacity allows filling of dissolution vessels for up to 3 dissolution tests (n = 6)
- Integration with DO probe facilitates individual vessel 2 and pilot DO levels for record and audit purposes
- Entire dissolution batch degassing ensures inter-vessel degassing consistency
- Volumetric dispensing with an accuracy better than + 1% as per USP, FDA and Ph. Eur. Requirements
- Special internal nozzle design allows the degassing of surfactant and frothing media like Sodium Lauryl Sulphate (SLS) and avoid frothing Tween
- Programmable deaeration time, temperature and vacuum for effective deaeration of different media
- Maintains media under vacuum at desired temperature while in idle condition
- Fast aspiration, heating and degassing allows 8 litres of media to be dispensed in less than 15 mins
- Fast filling of dissolution vessels at a dispensing rate of 1000 ml/min
- Programmable rinse and cleaning cycle allows the instrument to run through a prevalidated self cleaning procedure
- On-board memory stores 10 batch information of degassed media
- 360 turning radius allows it to be maneuvered in tight places
- Multifunction single hose for dispensing, aspirating, cleaning, rinsing, and draining ensures no lifting of heavy media tanks
- Print report consisting validation, calibration and main cylce report . Main cycle report consists set vacuum level, dispensed volume, degassing temperature ,O2 level, O level of pilot reading , O level of individual vessel and maximum vacuum level attained
Embedded printer for printing reports
Features of optical DO probe:
- Compliance with ASTM International Method D 888 - 05
- Quick response optical sensor technology
- No membrane : Cost-effective optical sensor technologyallows maintenance – free operation
- No electrlyte, no drift
- Real time temperature compensation
- Provides reliable and reproducible results
- Rugged and robust stainless steel construction
Specifications :
Model | EMD-21PDO |
Accuracy | +/-1% |
Flow rate | 1000 ml/min. (During run cycle) |
Dispense Volume | Programmable from 0.2 L to 4 L |
Temperature Range | 20° C to 50° C |
Temperature Accuracy | ±0.5C |
Media preparation time | 25 min |
O2 level | Normal tap water which contains 85% of dissolved O before degassing will contain 40% of dissolved O2 after degassing |
Hose | Single |
Media tank | S.S. 316, Teflon coated |
Vacuum pump | 760 mmHg |
Piston pump | Ceramic metering piston pump |
Accuracy | ±1% |
Power | 1200 watts |
Power supply | 230V AC/50 Hz. +10% |
Humidity | 70% max. |
Storage temperature | 10° C to 50° C |
Operating temperature | 10° C to 30° C |
Dimensions | ǿ 55 x 92 H cm |
Weight | Net 100 kgs Gross 132 kgs |