Construction :
- Constructed in double wall-PUF Insulated.
- Interior : Stainless Steel 22 G
- Exterior : G CRCA MS Sheet 1.2 mm
- Doubled walled typed with transparent full view Tufund glass door.
- Door : fitted on the chamber by heavy-duty hinges.
- Locking arrangement.
- Gaskets on the door. Fitted with Castor lockable wheels for ease of mobility
- Full front opening door.
Specifications :
Capacity | 20 Standard Blood Bags of each 450 ml capacity | ||
Digital Temp. Controller | Microprocessor Based Digital Temperature Controller cum Indicator used for controlling 4o± 1o C, accuracy : 0.1o C, Fan Fitted in the Inner side for proper air circulation in chamber | ||
Air circulation | Having provision for Air Circulation through internal fan facility. Temp gradient less than 0.5° C | ||
Inspection Lamp | One low watt bulb for internal inspection automatically / manual operated, | ||
Power Supply | 230Volts, Single Phase 50HZ, | ||
Trays / Shelves | No of Trays 3 in S.S. for proper Storage of blood bags. It is fitted on Imported Telescopic Channels. | ||
Safety | ○ Electronic safety device. ○ Low & High Voltageprotector. ○ Audiovisual alarming case of power failure,opening the Door, & in case of high & Low internal temperature of BBR. | ||
Optional Accessories | 1. Weekly Circular Chart Recorder:- For Continuous Temp. Recording Weekly Temperature Circular Chart Recorder (Range -10oC to + 10oC), Accuracy: 0.2oC, for Temp. Display visually & for continuous temperature monitoring. 2. Servo Voltage Stabilizer | ||