Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Auto Karl Fischer Titrimeter

Auto Karl Fischer Titrimeter

Product Details :

Temperature40 Deg C
BrandManti lab Solutions
ModelMT-139
Application


Auto Karl Fischer Titrimeter is an automatic apparatus designed to carry out Karl Fischer Titrations for accurate determination of moisture content in various samples. The instrument comprises of two units i.e CONTOL UNIT TITRATION UNIT. The CONTROL UNIT controls the complete automatic titration. On pressing the START button, the titration starts and the KF reagent is added automatically from the burette, through a solenoid valve into the titration vessel. As soon as the end point is reached, the titration is stopped automatically by the Control Unit. It has fully solid state circuitry incorporating advanced IC technology.

Titration unit contains the magnetic stirring mechanism and the solenoid valve assembly. The cabinet is designed to serve as a titration stand.

Karl Fischer method for determination of moisture content is used in pharmaceutical and drug manufacturing industries, food stuff, sugar, wood, paper and textile industries, petroleum refineries, chemical and fertilizer units etc.


Features :

  • Auto Titration Control
  • Separate Control Titration Units
  • Auto Zero Burette
  • Adjustable Timer
  • Visual Audible Alarm


Specifications :


Control Unit
DisplayArray of 10 LEDs
TimerAdjustable Timer
BuzzerYes
Sample Size1-50 mg of Water
K.F.Dispensing Resolution0.05 ml.
Measuring ElectrodeDual Platinum
IndicationsRUN}
ADD } Display in Auto Mode
END}
DetectionCurrent sensing by electronic control
DimensionsL340 X B250 X H125 mm. (Approx.)
Weight2 kg. (Approx.)
Titration Unit
Titration CellAir Tight 200ml.
StirrerVariable speed magnetic stirrer with On/Off control
K.F.FlowAutomatic Sensitive Control Valve for KF Reagent
BurettePressure filling auto zero burette with 0.05 ml resolution
Power230V ±10% 50Hz, AC
DimensionsL325 X B180 X H98 mm (Approx.)
Weight3 Kg. (Approx.)
AccessoriesBurette (Clear Glass) 10 ml * Reservoir Bottle 500 ml
Moisture trap with rubber bellow for air pumping
Titration Cell
Dual Platinum Electrode
Solenoid Valve Assembly
Magnetic Capsule
Set of 2 Clamps
Rubber Tubing
Nozzle
Instruction Manual

Note : Specifications are subject to change due to continuous improvements.

Write a review

Note: HTML is not translated!
    Bad           Good

Auto Karl Fischer Titrimeter

  • ₹0

Seller information

Please sign in to contact Manti Lab Solutions

Tags: Fischer Titrator