- Pharmaceutical Industry
- Environment Monitoring
- Commodity Inspection
- Food Inspection
- Agricultural Chemistry
- Teaching in Colleges & Universities
- Mettallurgy
- Geology
- Machine Manufacturing
- Petrochemical Industries
- Water and waste water Labs
- Food and Beverages Labs
Features:
- Unique optical system, 1200 lines/mm grating and imported receiver to ensure its sound performance
- The instrument is supply with PC software expands the capabilities of the instrument includes Photometric analysis, Quantitative analysis, Spectral scanning, Kinetics, DNA/protein analysis,Multi-wavelength test, etc.
- Adopts 51 microcomputer system, and store 200 test data and 100 standard curves;
- Plug type deuterium lamp and tungsten lamp transfer lamps without optics debugging.
STANDARD CONFIGURATION
Main Set | 1 no. |
USB Cable Accessory kit, Including a 4 glass cell (1cm) | 1 no. |
2 Quartz Cell (1 cm) | 1 no. |
Operating Manual Software Manual | 1 no. |
Dust Cover | 1 no. |
Software CD | 1 no. |
Software Key | 1 no. |
Power Cable | 1 no. |
Model | LJ-2371 |
Optical System | Single Beam, Grating 1200 lines/mm |
Wavelength Range | 190-1100nm |
Spectral Bandwidth | 2 nm |
Wavelength Accuracy | 1nm |
Wavelength Repeatability | <=0.2 nm |
Photometric Accuracy | 0.5%T |
Photometric Repeatability | 0.2%T |
Photometric Range | 0 - 200%T, -0.3 - 3A |
Stray light | 0.15%T |
Stability | +0.0002A/h @500nm |
Baseline Flatness | +0.002 A |
Noise | +0.001 A |
Wavelength Setting | Auto |
Light Source | Imported Deuterium & Tungsten lamp |
Display | 128*64 LCD |
Detector | Imported Silicon Photodiode |
Output | USB port & Parallel port (Printer) |
Power | AC 220V/50Hz; AC 110V/60Hz |
Dimension | 590*460*220mm |
Net weight | 16 kg |
Shipping Size | 700*580*340mm |
Weight | 20kg |