Our website uses cookies. They help us understand how customers use our website so we can give you the best experience possible. By continuing to browse this site or choosing to close this message, you give consent for cookies to be used. The cookies are not used for advertising. This applies to visitors from EU.
Process Calibrators
Process Calibrators
Product Details:

• Measures and simulates RDT, mV, mA signals simultaneously
• Accuracy @ 0.05% F. S
• 8 Bit Microprocessor, allows user to perform multiple functions in parallel
• Large, easy to read LCD display
• Industrial standard binding post terminals for easy field wiring
• True one-hand operation
• Universal power supply of 90 to 270 V AC or internal rechargeable battery operation

Process Calibrator
enquiry image
Description :

SYSCAL series of Calibrators provide features & functionality which is not found in other Calibration equipment. Designed for use in today’s demanding process industries, the CAL-001 is ideal for use in the field or the laboratory. The CAL-001 is more than a calibrator, it is also a high precision field diagnostic tool for troubleshooting your most difficult process applications.
Technical Specifications :
Modes of Operation : Source & Sink
Signals for operations:
• RTD ( Pt 100, Pt 200, CU 53 )
• TC ( J, K, R, S, T, B, E types )
• mV ( -200 to +200 mV for sink mode & -100 to +100 mV for source mode )
• mA (0 to 24 mA )
• V ( 0 to 10 V )
Common mode rejection
>120 dB at AC operation
Normal mode rejection
>60 dB at 50 or 60 Hz
Temperature Stability
± 0.01% peroC
Storage temperature range
-10 to 60oC
Input-Output isolation
500 V AC/DC
Measurement sampling time
250 ms
Display
LCD, 16 x 2 display with backlit
Accuracy
± 0.05% overall for all parameters
Linearity
0.05%
Response Time
<100 msec
Functionality
Auto / Manual mode, Stepwise increment / decrement, min-max step for calibration
Power Supply
External AC Supply ( universal 90 to 270 V AC )
OR Internal rechargeable Ni-cd Battery ( 12 V, 1.2 Ah ) with charger
( Operational time with Battery 6 hrs. & recharging time 12 hrs. upto 90% )
Load
600 Ohm for 0 to 24 mA, 4.7 Kohm & above for 0 to 10 V DC
Transmitter Supply
24 V DC, 30 mA
Case
Injection molded ABS with internal metal RFI coating
Dimension
290 (L) x 200 (W) x 110 (H) in mm ( DIN std. )
Weight
Net 4 kg & with charger 4.5 kg
Mounting
Table top Mounting
Accessories
Resin carrying case, Banana Cables, Mains Chord
Operating Conditions
Temperature 0 to 55oC / Humidity 5 to 95% non condensing