Fiber Extraction FIBERTEC 8000
Fully automatic -Crude Fiber estimation system
Application : Fibre determinations according to Weende, van Soest, etc. Single or sequential extractions
including
boiling, using externally preheated reagents, rinsing and filtration are performed under reproducible
and controlled conditions. Should be able to determine Crude Fibre (acc. to Weende). • Neutral
Detergent Fibre.• Acid Detergent Fibre. • Acid Detergent Lignin.
Six samples should be processed simultaneously.
Accessories for batch handling should be provided .
Machine should have Integral extraction and filtration.
No sample transfer and no loss of sample.
Machine should automatically heat and add all the reagents .
Overnight unattended operation should be possible.
User involvement for extraction should not be more then 3 minutes with the machine.
Separate unit for solvent dehydration, lignin determination and defatting.
Official approval: ISO 6865, 92/89 EEC and AOAC 2002.04 (feed).
Extensive application service and application notes should be provided whenever required.
Few application notes following AOAC methods for different samples should be enclosed with the
bid
Sample size: 0.5 - 3 g
Measuring range: 0.1 - 100%
Repeatability: < 1% relative at 5 - 30 % fibre level
Accuracy: According to official methods
Capacity per batch: 6 samples simultaneously
Capacity per day: Up to 36 analyses (Crude Fibre method acc. to Weende) per day
Hot Extraction Unit comprising: Hot Extractor, Reflector, Reagent Bottles, Hot Water Sprayer,
Beaker, Funnel, Water Suction Pump, Antifoaming Agent, Tubing, Holder for 6 crucibles, Stand for 6
crucibles, Crucibles (P2 standard, 2 sets of 6), Manual.
Cold Extraction Unit comprising: Cold Extractor, Spray Bottle, Tubing, T-tube, Spare Part Kit, Stand
for 6 crucibles, Crucibles (P2 standard 40 - 100 μm, set of 6), Antifoaming Agent (octanol), Celite, 1 l
“Sample preparation mill : sample preparation mill with cyclone technique should be supplied from the
same manufacturer for preparation of homogeneous sample of 0.5 mm,1 mm ,2mm size.sufficient
sieves and 50 collection bottles with caps should be supplied”