Click Image for Gallery
Reliable, Robots & User Friendly
Quick Features:
- Auto Air Purge between samples to avoid carry over.
- High Accuracy results with good repeatability at best Price.
- Auto Sleep mode for Lamp to prolong its life.
- 48 direct test access keys for user friendly operations.
- Dynamic temperature control by Peltier.
- Flow cell and cuvette reading Available.
- Real-time absorbance Curve and Temperature display.
- Inbuilt high quality Thermal Printer.
OPTICAL SYSTEM
- Flow Cell :- 30 pl quartz flow cell
- Light Source :- Halogen lamp 6V/10W
- Temperature :- 25‘C, 30°C, 37°C, RT
- Temperature Control :- By peltier device
- Photo Detector :- UV Enhanced Silicon photodiode
- Cuvette option :- 10mm square cuvette/ Round cuvette (optional)
- Filter :- 7 interference filters (340, 405,505,546,578,620&670nm) with 3 optional positions
- Filter Selection :- Stepper motor based filter wheel
MEASURING SYSTEM
- Analysis Method :- End point/ Fixed time/Kinetic/Turbidimetry /Absorbance
- Reaction Volume :- 350—l000u|
- Aspiration :- Peristaltic pump with stepper motor
- Photometric Range — 0.1 to 3.0 Absorbance
- Resolution :- 0001 Absorbance
- Repeatability :- Cv< 0.5%
- Carry over :- < 1%