METHODS OF ASSAY
- ABS/EP with Factor/KIN/Fix Time/Differential/ Multi point of 10 Standards.
OPTICAL SYSTEM
- Principle Absorption spectrometry
- Range of measurement From – 0.1 to 3.0A for 340nm, – 0.1 to 3.5A WL range 405 to 670nm
- Nominal spectral range Automatic via filter wheel
- Filters 340, 405, 450, 505, 546, 578, 630 and 670nm interference type
- Bandwidth 8 nm (± 2nm)
- Reading error ± 5% at 0.10 A
- Light source
- Quartz Halogen lamp 12 V / 20 W, 2000 hours
- Optional 6V/ 10 W
- Zero Automatic
- Settings Monochromatic and Bichromatic readings.
- Detector Schottkey type Photodiode - Silicon
THERMOSTAT SYSTEM
- Peltier regulated 25°, 30° & 37°C, RT programmed
- Accuracy ± 0.1°C
CUVETTES - DUAL MODE
Flow cell of 18 µL
Removable cuvettes Square micro plastic cuvettes & Round test tube diameter 13mm
PRINTER
- Type Fast Thermal Graphic
- Printing speed 50 mm/sec
- Dots per line 384 dots/line
- Paper Thermo sensitive, 57 mm wide
- RS232 For PC Connection
SCREEN AND KEYBOARD
- LCD blue LED back lit 240 x 128 pixels
- Keyboard Tactile Soft - Touch